Technical data
Trigger Commands 31
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 859
:TRIGger:TV:UDTV:PGTHan
Command :TRIGger:TV:UDTV:PGTHan <lower_limit>
The :TRIGger:TV:UDTV:PGTHan (Present Greater THan) command specifies
the minimum pulse width of the waveform used to arm the trigger used
during user- defined trigger mode.
<lower_limit> Minimum pulse width (time >), from 5 ns to 9.9999999 s.
Query
:TRIGger:TV:UDTV:PGTHan?
The query returns the currently selected minimum pulse width.
Returned Format
[:TRIGger:TV:UDTV:PGTHan] <lower_limit><NL>