Technical data

Trigger Commands 31
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 847
:TRIGger:TRANsition:SOURce
Command :TRIGger:TRANsition[{1 | 2}]:SOURce CHANnel<N>
The transition source command lets you find any edge in your waveform
that violates a rise time or fall time specification. The oscilloscope finds a
transition violation trigger by looking for any pulses in your waveform
with rising or falling edges that do not cross two voltage levels in the
amount of time you have specified.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<N> An integer, 1 - 4.
Query
:TRIGger:TRANsition:SOURce?
The query returns the currently defined transition source for the trigger
transition violation.
Returned Format
[:TRIGger:TRANsition:SOURce] CHANnel<N><NL>