:TRIGger:SHOLd:SetupTIMe
Command :TRIGger:SHOLd[{1 | 2}]:SetupTIMe <time>
This command specifies the amount of setup time used to test for both a
setup and hold trigger violation. The setup time is the amount of time that
the data must be stable and valid before a clock edge.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<time> Setup time, in seconds.
Query
:TRIGger:SHOLd:SetupTIMe?
The query returns the currently defined setup time for the setup and hold
trigger violation.
Returned Format
[:TRIGger:SHOLd:SetupTIMe] <time><NL>