Technical data
Trigger Commands 31
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 823
:TRIGger:SEQuence:RESet:ENABle
Command :TRIGger:SEQuence:RESet:ENABle {{ON | 1} | {OFF | 0}}
This command turns the Reset feature on or off for the sequential trigger.
The Reset feature allows you to specify a length of time such that if this
time is exceeded between when the TERM1 event occurs and when the
TERM2 event occurs, the sequential trigger is reset and the oscilloscope
returns to looking for the TERM1 event without triggering. If the Delay
feature (remote command :WAIT) is used as well then the Reset timer does
not start counting down until after the delay period is complete.
You can also base the Reset condition on a logical pattern. If the specified
pattern is found between when the TERM1 occurs and the TERM2 event
occurs, the sequential trigger resets and goes back to looking for the
TERM1 event without triggering. The delay feature does not impact a
logical pattern Reset as the pattern is searched for immediately after the
TERM1 event occurs regardless of whether of not the Delay period is
complete.
If the Reset feature is enabled, the AND qualifier cannot be used for the
TERM1 state.
Query
:TRIGger:SEQuence:RESet:ENABle?
The query returns whether or not the Reset feature is enabled.