Technical data

814 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
31 Trigger Commands
:TRIGger:PWIDth:WIDTh
Command :TRIGger:PWIDth[{1 | 2}]:WIDTh <width>
This command specifies how wide a pulse must be to trigger the
oscilloscope.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<width> Pulse width, which can range from 250 ps to 10 s.
Query
:TRIGger:PWIDth:WIDTh?
The query returns the currently defined width for the pulse.
Returned Format
[:TRIGger:PWIDth:WIDTh] <width><NL>