:TRIGger:PATTern:CONDition
Command :TRIGger:PATTern[{1 | 2}]:CONDition {ENTered | EXITed
| {GT,<time>[,PEXits|TIMeout]}
| {LT,<time>}
| {RANGe,<gt_time>,<lt_time>}}
This command describes the condition applied to the trigger pattern to
actually generate a trigger.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<gt_time> The minimum time (greater than time) for the trigger pattern, from 10 ns
to 9.999 s.
<lt_time> The maximum time (less than time) for the trigger pattern, from 10.5 ps to
10 s.
<time> The time condition, in seconds, for the pattern trigger, from 100 ps to
10 s.
When using the GT (Present >) parameter, the PEXits (Pattern Exits) or
the TIMeout parameter controls when the trigger is generated.
Query
:TRIGger:PATTern:CONDition?
The query returns the currently defined trigger condition.
Returned Format
[:TRIGger:PATTern:CONDition] {ENTered|EXITed
| {GT,<time>[,PEXits|TIMeout]}
| {LT,<time>}
| {RANGe,<gt_time>, <lt_time>}}<NL>