Technical data

Trigger Commands 31
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 803
:TRIGger:GLITch:POLarity
Command :TRIGger:GLITch[{1 | 2}]:POLarity {POSitive | NEGative}
This command defines the polarity of the glitch as positive or negative.
The trigger source must be set using the :TRIGger:GLITch:SOURce
command. The optional [{1 | 2}] parameter sets whether the trigger mode
goes with the TERM1 or TERM2 state if sequential triggering is being used.
Query
:TRIGger:GLITch:POLarity?
The query returns the currently selected glitch polarity.
Returned Format
[:TRIGger:GLITch:POLarity] {POSitive | NEGative}<NL>