Technical data

Trigger Commands 31
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 801
:TRIGger:EDGE:SOURce
Command :TRIGger:EDGE[{1 | 2}]:SOURce {CHANnel<N> | AUX | LINE}
The :TRIGger:EDGE:SOURce command selects the source for edge mode
triggering. This is the source that will be used for subsequent
:TRIGger:EDGE:SLOPe commands or queries.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<N> An integer, 1 - 4.
Query
:TRIGger:EDGE:SOURce?
The query returns the currently selected edge mode trigger source.
Returned Format
[:TRIGger:EDGE:SOURce] {CHANnel<N> | AUX | LINE}<NL>