Technical data
792 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
31 Trigger Commands
:TRIGger:DELay:EDELay:COUNt
Command :TRIGger:DELay:EDELay:COUNt <edge_number>
This command sets the event count for a Delay By Event trigger event.
<edge_num> An integer from 0 to 16,000,000 specifying the number of edges to delay.
Query
:TRIGger:DELay:EDELay:COUNt?
The query returns the currently defined number of events to delay before
triggering on the next Trigger On condition in the Delay By Event trigger
mode.
Returned Format
[:TRIGger:DELay:EDELay:COUNt] <edge_number><NL>