31 Trigger Commands
:TRIGger:SWEep
Command :TRIGger:SWEep {AUTO | TRIGgered | SINGle}
The :TRIGger:SWEep command selects the oscilloscope sweep mode. New
programs should use :RUN and :SINGle for run control and this command
for AUTO and TRIGgered for sweep control. The SINGle sweep control
should not be used.
AUTO When you select AUTO, if a trigger event does not occur within a time
determined by the oscilloscope settings, the oscilloscope automatically
forces a trigger which causes the oscilloscope to sweep. If the frequency of
your waveform is 50 Hz or less, you should not use the AUTO sweep mode
because it is possible that the oscilloscope will automatically trigger before
your waveform trigger occurs.
TRIGgered When you select TRIGgered, if no trigger occurs, the oscilloscope will not
sweep, and the previously acquired data will remain on the screen.
SINGle When you select SINGle, if no trigger occurs, the oscilloscope will not
sweep, and the previously acquired data will remain on the screen. Do not
use in new programs.
Query
:TRIGger:SWEep?
The query returns the specified channel's trigger level.
Returned Format
[:TRIGger:SWEep] {AUTO | TRIGgered}<NL>