Technical data

Trigger Commands 31
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 779
:TRIGger:LTHReshold
Command :TRIGger:LTHReshold CHANnel<N>,<level>
This command specifies the low threshold voltage level for the selected
trigger source. This command specifies the low threshold voltage level for
the selected trigger source. Set the low threshold level to a value
considered to be a low level for your logic family; your data book gives
two values, V
IL
and V
OL
.
<N> An integer, 1 - 4.
<level> A real number for the voltage level for the trigger source.
Query
:TRIGger:LTHReshold? CHANnel<N>
The query returns the currently defined low threshold for the trigger
source.
Returned Format
[:TRIGger:LTHReshold CHANnel<N>,] <level><NL>