Technical data

Trigger Commands 31
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 773
:TRIGger:AND:SOURce
Command :TRIGger:AND[{1 | 2}]:SOURce CHANnel<N>,{HIGH | LOW | DONTcare}
The :TRIGger:AND:SOURce command sets the logic value used to qualify
the trigger for the specified channel. The TRIGger:LEVel command
determines what voltage level is considered a HIGH or a LOW logic value.
If you set more than one channel to a HIGH or a LOW, then the multiple
channels are used to qualify the trigger.
The optional [{1 | 2}] parameter sets whether the AND qualifier goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<N> An integer, 1 - 4.
Query
:TRIGger:AND:SOURce? CHANnel<N>
The query returns the logic value for the designated channel.
Returned Format
[:TRIGger:AND:SOURce CHANnel<N>] {HIGH | LOW | DONTcare}<NL>