Technical data

720 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
28 Serial Data Equalization Commands
:SPRocessing:DFEQualizer:TAP:MIN
Command :SPRocessing:DFEQualizer:TAP:MIN <min_tap_value>
Some standards have upper and lower limits on the tap values. The
:DFEQualizer:TAP:MIN command sets the lower limit on taps determined
through optimization.
<min_tap_value
>
A real number
Example This example sets the Lower Limit field to 3.23.
myScope.WriteString ":SPRocessing:DFEQualizer:TAP:MIN 3.23"
Query :SPRocessing:DFEQualizer:TAP:MIN?
The :SPRocessing:DFEQualizer:TAP:MIN? query returns the Lower Limit
used in the DFE tap optimization.