Technical data
686 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
26 Serial Bus Commands
:SBUS<N>:SPI:CLOCk:TIMeout
Command :SBUS<N>:SPI:CLOCk:TIMeout <time_value>
<time_value> ::= time in seconds in NR3 format
The :SBUS<N>:SPI:CLOCk:TIMeout command sets the SPI signal clock
timeout resource in seconds from 100 ns to 10 s when the
:SBUS<N>:SPI:FRAMing command is set to TIMeout. The timer is used to
frame a signal by a clock timeout.
<N> An integer, 1 - 4.
Query
:SBUS<N>:SPI:CLOCk:TIMeout?
The :SBUS<N>:SPI:CLOCk:TIMeout? query returns current SPI clock
timeout setting.
Returned Format
[:SBUS<N>:SPI:CLOCk:TIMeout] <time value><NL>
See Also • ":SBUS<N>:SPI:CLOCk:SLOPe" on page 685
• ":SBUS<N>:SPI:SOURce:CLOCk" on page 688
• ":SBUS<N>:SPI:FRAMe:STATe" on page 687