Technical data
Root Level Commands 25
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 651
:AUToscale:VERTical
Command :AUToscale:VERTical {CHANnel<N>}
The :AUToscale:VERTical command autoscales the vertical position and
scaling for the corresponding channel without changing anything else (for
example, trigger or timebase settings).
Example This example automatically autoscales the vertical position and scale for
the waveform on Channel 1.
myScope.WriteString ":AUTOSCALE:VERTical CHAN1"
NOTE
If you are using software 2.10 or earlier, the command syntax is (lower-case "t" in
"vertical"):
AUToscale:VERtical <CHANnel 1 | CHANnel 2 | CHANnel 3 | CHANnel 4>