:MEASure:THResholds:ABSolute
Command :MEASure:THResholds:ABSolute <source>,
<upper_volts>,<middle_volts>,<lower_volts>
The :MEASure:THResholds:ABSolute command sets the upper level, middle
level, and lower level voltages that are used to calculate the measurements
that use them.
<source> {ALL | CHANnel<N> | FUNCtion<N> | WMEMory<N> | CLOCk | MTRend
| MSPectrum | EQUalized}
MTRend and MSPectrum sources are only available if the oscilloscope has
the EZJIT option installed and the feature is enabled.
The CLOCk source is only available if the oscilloscope has the High Speed
Serial option installed and the feature is enabled.
The EQUalized source is only available if the oscilloscope has the High
Speed Serial option and the Serial Data Equalization option installed and
the features are enabled. This command uses the Feed- Forward Equalized
(FFE) signal as the source. Setting the source to ALL does not affect the
individual channel settings which is the behavior as the user interface.
<N> is an integer, 1 - 4.
<upper_volts>
<middle_volts>
<lower_volts>
A real number specifying voltage thresholds.
Example This example sets the custom voltage thresholds to 0.9 volts for the upper
level, 0.5 volts for the middle level and 0.1 volts for the lower level on
channel 2.
myScope.WriteString ":SYSTEM:HEADER OFF" ' Response headers off.
myScope.WriteString ":MEASURE:THRESHOLDS:ABSOLUTE CHANNEL2,0.9,0.5,0.1"
Query :MEASure:THResholds:ABSolute? <source>
The :MEASure:THResholds:ABSolute? query returns the current settings for
upper level, middle level, and lower level voltages for the custom
thresholds.
Returned Format
[:MEASure:THResholds:ABSolute] <upper_volts>,<middle_volts>,<lower_volts
><NL>
Example This example returns the upper level, middle level, and lower level voltages
used to calculate the measurements on channel 1.
myScope.WriteString ":SYSTEM:HEADER OFF" ' Response headers off.
myScope.WriteString ":MEASURE:THRESHOLDS:ABSOLUTE? CHANNEL1"
strThresholds = myScope.ReadString
Debug.Print strThresholds