Technical data

566 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
24 Measure Commands
:MEASure:RESults?
Query :MEASure:RESults?
The :MEASure:RESults? query returns the results of the continuously
displayed measurements. The response to the MEASure:RESults? query is a
list of comma- separated values. If SENDvalid is ON, the results state is
returnedt.
If more than one measurement is running continuously, the values in the
:MEASure:RESults returned are duplicated for each continuous
measurement from the first to last (left to right) result displayed. Each
result returned is separated from the previous result by a comma. There is
a maximum of five continuous measurements that can be continuously
displayed at a time.
Returned Format
[:MEASure:RESults] <result_list><NL>
<result_list> A list of the measurement results separated with commas. The following
shows the order of values received for a single measurement if
:MEASure:STATistics is set to ON.
Min, max, mean, std dev, and # of meas are only returned if the
:MEASure:STATistics is ON. The result state is only returned if
:MEASure:SENDvalid is ON. See Table 23 for the meaning of the result
state codes.
If the :MEASure:STATistics is set to CURRENT, MAX, MEAN, MIN, or
STDDEV only that particular statistic value is returned for each
measurement that is on.
Example This example places the current results of the measurements in the string
variable, strResult, then prints the contents of the variable to the
computer's screen.
Dim strResult As String ' Dimension variable.
myScope.WriteString ":MEASURE:RESULTS?"
strResult = myScope.ReadString
Debug.Print strResult
Measure
ment
label
current result
state
min max mean std dev # of
meas