Technical data

Measure Commands 24
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 553
:MEASure:PERiod
Command :MEASure:PERiod [<source>[,<direction>]]
The :MEASure:PERiod command measures the period of the first complete
cycle on the screen using the mid- threshold levels of the waveform (50%
levels with standard measurements selected).
The source is specified with the :MEASure:SOURce command or with the
optional parameter following the :MEASure:PERiod command.
The algorithm is:
If the first edge on the screen is rising,
then
period = second rising edge time - first rising edge time
else
period = second falling edge time - first falling edge time
<source> {CHANnel<N> | FUNCtion<N> | WMEMory<N> | CLOCk | MTRend |
MSPectrum | EQUalized}
MTRend and MSPectrum sources are only available if the oscilloscope has
the EZJIT option installed and the feature is enabled.
The CLOCk source is only available if the oscilloscope has the High Speed
Serial option installed and the feature is enabled.
The EQUalized source is only available if the oscilloscope has the High
Speed Serial option and the Serial Data Equalization option installed and
the features are enabled. This command uses the Feed- Forward Equalized
(FFE) signal as the source.
<N> CHANnel<N> is an integer, 1 - 4.
FUNCtion<N> and WMEMory<N> are:
An integer, 1 - 4, representing the selected function or waveform memory.
<direction> {RISing | FALLing}
Specifies direction of edge to start measurement. When <direction> is
specified, the <source> parameter is required.
Using the <direction> parameter will set the "Measure All Edges" mode if it
is not currently set.
Example This example measures the period of the waveform.
myScope.WriteString ":MEASURE:PERIOD CHANNEL1"
Query :MEASure:PERiod? [<source>],<direction>