Technical data
Measure Commands 24
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 519
:MEASure:JITTer:SPECtrum:HORizontal:POSition
Command
:MEASure:JITTer:SPECtrum:HORizontal:POSition <position>
The :MEASure:JITTer:SPECtrum:HORizontal:POSition command sets the
jitter spectrum horizontal center frequency position.
<position> A real number for the center frequency position in Hertz.
Example This example sets the jitter spectrum horizontal center frequency position
to 250 kHz.
myScope.WriteString ":MEASURE:JITTER:SPECTRUM:HORIZONTAL:POSITION 250E3"
Query :MEASure:JITTer:SPECtrum:HORizontal:POSition?
The :MEASure:JITTer:SPECtrum:HORizontal:POSition? query returns the
current jitter spectrum horizontal center frequency position setting.
Returned format
[:MEASure:JITTer:SPECtrum:HORizontal:POSition] <value><NL>
<value> The jitter spectrum horizontal center frequency setting.
Example This example places the current setting of the jitter trend horizontal
center frequency position in the variable varValue, then prints the contents
of the variable to the computer's screen.
myScope.WriteString ":SYSTem:HEADer OFF"
myScope.WriteString ":MEASURE:JITTER:SPECTRUM:HORIZONTAL:POSITION?"
varValue = myScope.ReadNumber
Debug.Print FormatNumber(varValue, 0)
NOTE
This command is only available when the E2681A Jitter Analysis Software or the
N5400A/N5401A Software is installed.