Technical data
Measure Commands 24
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 511
:MEASure:HISTogram:RESolution
Command :MEASure:HISTogram:RESolution [<source>]
The :MEASure:HISTogram:RESolution command places the histogram bin
width measurement into the Measurements tab of the oscilloscope's user
interface.
The source is specified with the :MEASure:SOURce command or with the
optional parameter following the command.
The :MEASure:HISTogram commands only apply to Meas Histogram math
functions, the histogram waveform, or memories containing histograms.
<source> { FUNCtion<N> | WMEMory<N> | HISTogram}
<N> An integer, 1 - 4.
Example This example places into the Measurements tab the bin width of the
histogram stored in WMEMory1.
myScope.WriteString ":MEASure:HISTogram:RESolution WMEMory1"
Query :MEASure:HISTogram:RES? [<source>]
The :MEASure:HISTogram:RES? query returns the measurement of the bin
width of the histogram.
Returned Format
[:MEASure:HISTogram:RES]<value>[,<result_state>]<NL>
<value> The width of the histogram.
<result_state> If SENDVALID is ON, the result state is returned with the measurement
result. Refer to the MEASure:RESults command, for a list of the result
states.
Example This example returns the width of the current histogram and prints the
result to the computer's screen.
myScope.WriteString ":SYSTem:HEADer OFF" ' Response headers off.
myScope.WriteString ":MEASure:HISTogram:RESolution? WMEMory1"
varHistpp = myScope.ReadNumber
Debug.Print FormatNumber(varHistpp, 0)
See Also • ":FUNCtion<N>:MHIStogram" on page 314
• ":HISTogram:MODE" on page 338