Technical data

Measure Commands 24
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 509
:MEASure:HISTogram:PEAK
Command :MEASure:HISTogram:PEAK [<source>]
The :MEASure:HISTogram:PEAK command places the histogram number of
hits in the greatest peak measurement into the Measurements tab of the
oscilloscope's user interface.
The source is specified with the :MEASure:SOURce command or with the
optional parameter following the command.
The :MEASure:HISTogram commands only apply to Meas Histogram math
functions, the histogram waveform, or memories containing histograms.
<source> { FUNCtion<N> | WMEMory<N> | HISTogram}
<N> An integer, 1 - 4.
Example This example places into the Measurements tab the number of hits in the
greatest peak of the histogram stored in WMEMory1.
myScope.WriteString ":MEASure:HISTogram:PEAK WMEMory1"
Query :MEASure:HISTogram:PEAK? [<source>]
The :MEASure:HISTogram:PEAK? query returns the number of hits in the
greatest peak of the histogram measurement.
Returned Format
[:MEASure:HISTogram:PEAK]<value>[,<result_state>]<NL>
<value> The number of hits in the histogram peak.
<result_state> If SENDVALID is ON, the result state is returned with the measurement
result. Refer to the MEASure:RESults command, for a list of the result
states.
Example This example returns the number of hits in the greatest peak of the
current histogram and prints the result to the computer's screen.
myScope.WriteString ":SYSTem:HEADer OFF" ' Response headers off.
myScope.WriteString ":MEASure:HISTogram:PEAK? WMEMory1"
varHistpeak = myScope.ReadNumber
Debug.Print FormatNumber(varHistpeak, 0)
See Also ":FUNCtion<N>:MHIStogram" on page 314
":HISTogram:MODE" on page 338