Technical data
Measure Commands 24
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 503
:MEASure:HISTogram:M3S
Command :MEASure:HISTogram:M3S [<source>]
The :MEASure:HISTogram:M2S command places the histogram percentage
of points within two standard deviations of the mean measurement into
the Measurements tab of the oscilloscope's user interface.
The source is specified with the :MEASure:SOURce command or with the
optional parameter following the command.
The :MEASure:HISTogram commands only apply to Meas Histogram math
functions, the histogram waveform, or memories containing histograms.
<source> { FUNCtion<N> | WMEMory<N> | HISTogram}
<N> An integer, 1 - 4.
Example This example measures the percentage of points that are within three
standard deviations of the mean of the histogram.
myScope.WriteString ":MEASure:HISTogram:M3S HISTogram"
Query :MEASure:HISTogram:M3S? [<source>]
The :MEASure:HISTogram:M3S? query returns the measurement of the
percentage of points within three standard deviations of the mean of the
histogram.
Returned Format
[:MEASure:HISTogram:M3S]<value>[,<result_state>]<NL>
<value> The percentage of points within three standard deviations of the mean of
the histogram.
<result_state> If SENDVALID is ON, the result state is returned with the measurement
result. Refer to the MEASure:RESults command, for a list of the result
states.
Example This example returns the percentage of points within three standard
deviations of the mean of the current histogram and prints the result to
the computer's screen.
myScope.WriteString ":SYSTem:HEADer OFF" ' Response headers off.
myScope.WriteString ":MEASure:HISTogram:M3S? WMEMory1"
varHistm3s = myScope.ReadNumber
Debug.Print FormatNumber(varHistm3s, 0)
See Also • ":FUNCtion<N>:MHIStogram" on page 314
• ":HISTogram:MODE" on page 338