Technical data
Measure Commands 24
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 447
:MEASure:BPERiod
Command :MEASure:BPERiod <source>, <idle time>
The :MEASure:BPERiod command measures the time between the
beginning of a burst and the beginning of the next burst. The idle time is
the minimum time between bursts.
<source> {CHANnel<N> | FUNCtion<N> | WMEMory<N>}
<N> CHANnel<N> is an integer, 1 - 4.
FUNCtion<N> and WMEMory<N> are:
An integer, 1 - 4, representing the selected function or waveform memory.
<idle time> Minimum amount of idle time between bursts.
Example This example measures the burst period between two bursts on channel 4
(and with an idle time of 5 microseconds)
myScope.WriteString ":MEASURE:BPERiod CHAN4, 5e-6"
Query :MEASure:BPERiod? <source>, <idle time>
The :MEASure:BPERiod? query returns the burst period time.