Technical data

Mask Test Commands 23
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 425
:MTESt:SCALe:BIND
Command :MTESt:SCALe:BIND {{ON | 1} | {OFF | 0}}
The :MTESt:SCALe:BIND command enables or disables Bind 1 & 0 Levels
(Bind - 1 & 0 Levels for inverted masks) control. If the Bind 1 & 0 Levels
control is enabled, the 1 Level and the 0 Level controls track each other.
Adjusting either the 1 Level or the 0 Level control shifts the position of
the mask up or down without changing its size. If the Bind 1 & 0 Levels
control is disabled, adjusting either the 1 Level or the 0 Level control
changes the vertical height of the mask.
If the Bind - 1 & 0 Levels control is enabled, the - 1 Level and the 0 Level
controls track each other. Adjusting either the - 1 Level or the 0 Level
control shifts the position of the mask up or down without changing its
size. If the Bind - 1 & 0 Levels control is disabled, adjusting either the - 1
Level or the 0 Level control changes the vertical height of the mask.
Example This example enables the Bind 1 & 0 Levels control.
myScope.WriteString ":MTEST:SCALE:BIND ON"
Query :MTESt:SCALe:BIND?
The :MTESt:SCALe:BIND? query returns the value of the Bind 1&0 control
(Bind - 1&0 for inverted masks).
Returned Format
[:MTESt:SCALe:BIND?] {1 | 0}<NL>