Technical data
Mask Test Commands 23
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 421
:MTESt:NREGions?
Query :MTESt:NREGions?
The :MTESt:NREGions? query returns the number of regions that define
the mask.
Returned Format
[:MTESt:NREGions] <regions><NL>
<regions> An integer from 0 to 8.
Example This example returns the number of mask regions.
myScope.WriteString ":SYSTEM:HEADER OFF"
myScope.WriteString ":MTEST:NREGIONS?"
varRegions = myScope.ReadNumber
Debug.Print FormatNumber(varRegions, 0)