Technical data

418 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
23 Mask Test Commands
:MTESt:INVert
Command :MTESt:INVert {{ON | 1} | {OFF | 0}}
The :MTESt:INVert command inverts the mask for testing negative-going
pulses. The trigger level and mask offset are also adjusted. Not all masks
support negative- going pulse testing, and for these masks, the command is
ignored.
Example This example inverts the mask for testing negative- going pulses.
myScope.WriteString ":MTEST:INVERT ON"
Query :MTESt:INVert?
The :MTESt:INVert? query returns the current inversion setting.
Returned Format
[:MTESt:INVert] {1 | 0}<NL>