Technical data

386 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
22 Marker Commands
:MARKer:X2Position
Command :MARKer:X2Position <Bx_position>
The :MARKer:X2Position command sets the Bx marker position and moves
the Bx marker to the specified time with respect to the trigger time.
<Bx_position> A real number for the time at the Bx marker in seconds.
Example This example sets the Bx marker to 90 ns.
myScope.WriteString ":MARKER:X2POSITION 90E-9"
Query :MARKer:X2Position?
The :MARKer:X2Position? query returns the time at Bx marker in seconds.
Returned Format
[:MARKer:X2Position] <Bx_position><NL>
Example This example returns the current position of the Bx marker to the
numeric variable, varValue, then prints the contents of the variable to the
computer's screen.
myScope.WriteString ":SYSTEM:HEADER OFF" ' Response headers off.
myScope.WriteString ":MARKER:X2POSITION?"
varValue = myScope.ReadNumber
Debug.Print FormatNumber(varValue, 0)