Technical data

InfiniiScan (ISCan) Commands 19
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 349
:ISCan:MEASurement:FAIL
Command :ISCan:MEASurement:FAIL {INSide | OUTSide}
The :ISCan:MEASurement:FAIL command sets the fail condition for an
individual measurement. The conditions for a test failure are set on the
measurement selected by the :ISCan:MEASurement command.
When a measurement failure is detected by the limit test the oscilloscope
triggers and the trigger action is executed.
INSide INside causes the oscilloscope to fail a test when the measurement results
are within the parameters set by the :ISCan:MEASurement:LIMit and
:ISCan:MEASurement:ULIMit commands.
OUTSide OUTside causes the oscilloscope to fail a test when the measurement
results exceed the parameters set by the :ISCan:MEASurement:LLIMit and
the :ISCan:MEASurement:ULIMit commands.
Example The following example causes the oscilloscope to trigger when the
measurements are outside the lower or upper limits.
myScope.WriteString ":ISCAN:MEASUREMENT:FAIL OUTSIDE"
Query :ISCan:MEASurement:FAIL?
The query returns the current set fail condition.
Returned Format
[:ISCan:MEASurement:FAIL] {INSide | OUTSide}<NL>
Example The following example returns the current fail condition and prints the
result to the controller's screen.
Dim strFAIL As String
myScope.WriteString ":ISCAN:MEASUREMENT:FAIL?"
strFAIL = myScope.ReadString
Debug.Print strFAIL