Technical data

Channel Commands 12
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 217
:CHANnel<N>:PROBe:PRECprobe:CALibration
Command :CHANnel<N>:PROBe:PRECprobe:CALibration <cal_string>
The :CHANnel<N>:PROBe:PRECprobe:CALibration command specifies the
name of the PrecisionProbe or PrecisionCable calibration to use for the
specified channel and probe.
<N> An integer, 1 - 4.
<cal_string> A quoted string that is the name of the PrecisionProbe or Precision Cable
calibration.
Example This example says to use the PrecisionProbe or PrecisionCable calibration
named "2- 8- 2" for channel 1.
myScope.WriteString ":CHANNEL1:PROBE:PRECprobe:CALibration "2-8-2""
Query :CHANnel<N>:PROBe:PRECprobe:CALibration?
The :CHANnel<N>:PROBe:PRECprobe:CALibration? query returns the
currently specified name for the selected channel's PrecisionProbe or
PrecisionCable calibration.
Returned Format
[:CHANnel<N>:PROBe:PRECprobe:CALibration] <cal_string><NL>
See Also ":CHANnel<N>:PROBe:ACCAL" on page 198
":CHANnel<N>:PROBe:PRECprobe:MODE" on page 218
":CHANnel<N>:PROBe:PRECprobe:ZSRC" on page 219
":CHANnel<N>:PROBe:PRECprobe:BANDwidth" on page 216