12 Channel Commands
:CHANnel<N>:PROBe:PRECprobe:BANDwidth
Command :CHANnel<N>:PROBe:PRECprobe:BANDwidth {AUTO | {MANual, <bandwidth>}
| {BOOSt, <boost_dB>}}
The :CHANnel<N>:PROBe:PRECprobe:BANDwidth command specifies how
the limit of PrecisionProbe or PrecisionCable correction/boosting is
determined.
<N> An integer, 1 - 4.
AUTO PrecisionProbe or PrecisionCable normally sets the bandwidth to a value
that has a small amount of boosting in the frequency response.
MANual,
<bandwidth>
Let you manually specify a bandwidth limit at which to stop applying
correction.
BOOSt,
<boost_dB>
Lets you specify a dB limit at which to stop applying correction.
Example This example specifies that, for PrecisionProbe or PrecisionCable on
channel 1, correction/boosting should stop being applied at a 3 dB limit.
myScope.WriteString ":CHANNEL1:PROBE:PRECprobe:BANDwidth BOOSt, 3"
Query :CHANnel<N>:PROBe:PRECprobe:BANDwidth?
The :CHANnel<N>:PROBe:PRECprobe:BANDwidth? query returns the
current PrecisionProbe or PrecisionCable corrected bandwidth setting for
the selected channel.
Returned Format
[:CHANnel<N>:PROBe:PRECprobe:BANDwidth] {AUTO | {MANual, <bandwidth>}
| {BOOSt, <boost_dB>}}<NL>
See Also ":CHANnel<N>:PROBe:ACCAL" on page 198
":CHANnel<N>:PROBe:PRECprobe:MODE" on page 218
":CHANnel<N>:PROBe:PRECprobe:CALibration" on page 217
":CHANnel<N>:PROBe:PRECprobe:ZSRC" on page 219