Technical data

152 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
9 Acquire Commands
Interaction
between
:ACQuire:SRATe
and
:ACQuire:POINts
If you assign a sample rate value with :ACQuire:SRATe or a points value
using :ACQuire:POINts the following interactions will occur. "Manual"
means you are setting a non- AUTO value for SRATe or POINts.
Example This example sets the memory depth to 500 points.
myScope.WriteString ":ACQUIRE:POINTS 500"
Table 10 50 M Memory Option Installed
Sampling mode and sample rate Trigger Sweep
Single Auto or Triggered
Normal, Peak Detect (full sample rate),
and High Resolution modes
16 to 51,250,000
Normal and High Resolution modes
with averaging
16 to 2,050,000
Peak Detect Mode at < full sample
rate
16 to 51,250,000
Equivalent Time Mode 16 to 262,144
Table 11 20 M Memory Option Installed
Sampling mode and sample rate Trigger Sweep
Single Auto or Triggered
Normal, Peak Detect (full sample rate),
and High Resolution modes
16 to 20,500,000
Normal and High Resolution modes
with averaging
16 to 2,050,000
Peak Detect Mode at < full sample
rate
16 to 20,500,000
Equivalent Time Mode 16 to 262,144
SRATe POINts Result
AUTO Manual POINts value takes
precedence (sample rate is
limited)
Manual AUTO SRATe value takes precedence
(memory depth is limited)
Manual Manual SRATe value takes precedence
(memory depth is limited)