Technical data
150 Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
9 Acquire Commands
:ACQuire:POINts
(Memory depth)
Command
:ACQuire:POINts {AUTO | <points_value>}
The :ACQuire:POINts command sets the requested memory depth for an
acquisition. Before you download data from the oscilloscope to your
computer, always query the points value with the :WAVeform:POINts?
query or :WAVeform:PREamble? query to determine the actual number of
acquired points.
You can set the points value to AUTO, which allows the oscilloscope to
select the optimum memory depth and display update rate.
<points_value> An integer representing the memory depth.
The range of points available for a channel depends on the oscilloscope
settings of sampling mode, sampling rate, and trigger sweep. The following
tables show the range of memory values for the different memory options.
Table 6 1 G Memory Option Installed
Sampling mode and sample rate Trigger Sweep
Single Auto or Triggered
Normal, Peak Detect (full sample rate),
and High Resolution (full sample rate)
Modes
16 to 1,049,600,000
Normal and High Resolution modes
with averaging
16 to 2,050,000
High Resolution Mode at < full sample
rate
16 to 524,800,000 16 to 262,400,000
Peak Detect Mode at < full sample
rate
16 to 262,400,000 16 to 131,200,000
Equivalent Time Mode 16 to 262,144
Table 7 500 M Memory Option Installed
Sampling mode and sample rate Trigger Sweep
Single Auto or Triggered
Normal, Peak Detect (full sample rate),
and High Resolution modes
16 to 512,500,000
Normal and High Resolution modes
with averaging
16 to 2,050,000