Technical data

Reference 36
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 1071
CSV and TSV Header Format
Revision Always 0 (zero).
Typ e How the waveform was acquired: normal, raw, interpolate, average, or
versus. When this field is read back into the scope, all modes, except
versus, are converted to raw. The default value is raw.
Start Starting point in the waveform of the first data point in the file. This is
usually zero.
Points The number of points in the waveform record. The number of points is set
by the Memory Depth control. The default value is 1.
Count or
Segments
For count, it is the number of hits at each time bucket in the waveform
record when the waveform was created using an acquisition mode like
averaging. For example, when averaging, a count of four would mean every
waveform data point in the waveform record has been averaged at least
four times. Count is ignored when it is read back into the scope. The
default value is 0.
Segments is used instead of Count when the data is acquired using the
Segmented acquisition mode. This number is the total number of segments
that were acquired.
XDispRange The number of X display range columns (n) depends on the number of
sources being stored. The X display range is the X-axis duration of the
waveform that is displayed. For time domain waveforms, it is the duration
of time across the display. If the value is zero then no data has been
acquired.
XDispOrg The number of X display origin columns (n) depends on the number of
sources being stored. The X display origin is the X- axis value at the left
edge of the display. For time domain waveforms, it is the time at the start
of the display. This value is treated as a double precision 64- bit floating
point number. If the value is zero then no data has been acquired.
XInc The number of X increment columns (n) depends on the number of
sources being store. The X increment is the duration between data points
on the X axis. For time domain waveforms, this is the time between
points. If the value is zero then no data has been acquired.
XOrg The number of X origin columns (n) depends on the number of sources
being store. The X origin is the X-axis value of the first data point in the
data record. For time domain waveforms, it is the time of the first point.
This value is treated as a double precision 64- bit floating point number. If
the value is zero then no data has been acquired.
XUnits The number of X units columns (n) depends on the number of sources
being store. The X units is the unit of measure for each time value of the
acquired data.