Technical data

Status Reporting 6
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference 107
Operation Status Register
This register hosts the following bits:
ACQ DONE bit 0
PROC DONE bit 1
WAIT TRIG bit 5
MASK bit 9
AUTO TRIG bit 11
OVLR bit 12
The ACQ DONE done bit is set by the Acquisition Done Event Register.
The PROC DONE bit is set by the Process Done Event Register and
indicates that all functions and all math processes are done.
The WAIT TRIG bit is set by the Trigger Armed Event Register and
indicates the trigger is armed.
The MASK bit is set whenever at least one of the Mask Test Event Register
bits is enabled.
The AUTO TRIG bit is set by the Auto Trigger Event Register.
The OVLR bit is set whenever at least one of the Overload Event Register
bits is enabled.
If any of these bits are set, the OPER bit (bit 7) of the Status Byte
Register is set. The Operation Status Register is read and cleared with the
OPER? query. The register output is enabled or disabled using the mask
value supplied with the OPEE command.