7
Remove the three M4 screws located on the top edge of the rear panel of the
oscilloscope (torque to 18 in- lbs when installing), then turn the instrument on
its side and remove the ten M3 screws located on the underside of the
oscilloscope that attach the outer cover to the chassis (torque to 5 in- lbs when
installing).
Figure 5-3
Figure 5-4
8
Carefully slide the cover toward the rear of the instrument and off the chassis.