Specifications

Chapter 4: Troubleshooting
Power Supply Troubleshooting
57
plugged into the LO or COMMON input of the DVM. On a good backplane assembly,
you should measure about 700 ohms. If there is a short on this board, you would
measure something very near 0 ohms.
3 If there is a short, disconnect the power cable to the front panel (see Figure 4- 9) and
redo the resistance measurement.
4 If the short is gone, replace the front panel assembly and verify that the oscilloscope
now powers on.
5 If there is still a short circuit, disconnect the two large connectors that provide power
to the motherboard (J1201 and J1202) from the backplane assembly (see the circled
cable connections in Figure 4- 11) and redo the resistance measurements across the
FET (Q1409) or the bulk supply input pins on the backplane assembly.
6 If the short is gone, replace the motherboard assembly and confirm that the unit will
power on.
Figure 4-10 Power supply disconnect
Figure 4-11 Cable connection into backplane