Specifications

Chapter 3: Testing Performance
Analog Bandwidth -- Maximum Frequency Test
32
c Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
d Select Setup > Acquisition and then set up the acquisition parameters as follows:
Sin(x)/x Interpolation = Auto
Analog Averaging = Disabled
Analog Memory Depth = Automatic
Analog Sampling rate = 80 GSa/s (maximum)
e Select Measure > Voltage > V rms.
Click here and enter 16 ns