Specifications

Chapter 3: Testing Performance
Analog Bandwidth -- Maximum Frequency Test
30
Analog Bandwidth - - Maximum Frequency Test
CAUTION Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
Specification
Equipment Required
Let the oscilloscope warm up before testing.
The oscilloscope under test must be warmed up (with the oscilloscope application running) for at least
30 minutes prior to the start of any performance test.
Analog Bandwidth (-3 dB)
MSO/DSO/DSAX91304A 13.0 GHz
MSO/DSO/DSAX91604A 16.0 GHz
MSO/DSO/DSAX92004A 20.0 GHz
MSO/DSO/DSAX92504A 25.0 GHz
MSO/DSO/DSAX92804A 28.0 GHz
DSO/DSA93004L 30.0 GHz
MSO/DSO/DSAX93204A 32.0 GHz
Description Critical Specifications Recommended Model/
Part Numbers
Microwave CW
Generator
Maximum Frequency ≥ 32 GHz
Power range: -20 dBm to +16 dBm into 50Ω
Output resistance = 50Ω
Agilent E8257D with Opt 540
Power Splitter 2 Resistor Power Splitter
Max Frequency 32 GHz
Agilent 11667C
Power Meter Agilent E-series with power sensor compatibility Agilent E4418B or E4419B
Power Sensor Maximum Frequency 32 GHz
Power range: -24 dBm to +16 dBm
Agilent 8487A or Agilent E4413B
with 8485A-033
Microwave Cable 50Ω Characteristic Impedance
2.4 mm (m) to 2.4 mm (m) SMA connectors
Max Frequency 32 GHz
Agilent N5180-60204
Adapter 2.4 mm (m) to 3.5 mm (m) (qty. 1) Agilent 11901A
Adapter 2.4 mm (m) to 3.5 mm (f) (qty. 2) Agilent 11901C
Connector Saver 3.5 mm (f)-(f), shipped with MSO/DSO/DSAX 90000A
and DSO/DSA93004L models
Agilent 5061-5311 (used with 20,
25, 28, 30, and 32 GHz oscilloscope
models)
Agilent 1250-3758 (used with 13
and 16 GHz oscilloscope models)