Specifications

Chapter 3: Testing Performance
DC Gain Accuracy Test
28
5 Set the channel 1 vertical scale value to 10 mV/div by using the front panel control or:
a Pull down the Setup menu and select Channel 1 or click the Channel 1 setup icon.
b Change the vertical scale of channel 1 to 10 mV/div.
c Select the Vavg measurement as shown below.
d When the Enter Measurement Info window appears, ensure that the V avg function
is set up as follows and then click OK:
Source = Channel 1
Measurement Area = Entire Display
6 Press [Clear Display] on the scope, wait for the #Avgs value (top right corner of screen)
to return to 256, and then record the scope's mean V avg reading in the DC Gain
Accuracy Test section of the Performance Test Record.
Notes
For all scope readings in this procedure, use the mean value in the Measurements display area at the
bottom of the screen.
If a question mark appears in front of any of the values at the bottom of the screen, press [Clear Display]
on the scope, wait for the #Avgs value to return to 256, and then record the scope reading.
V avg
measurement