Specifications
Chapter 3: Testing Performance
DC Gain Accuracy Test
27
b When the Acquisition Setup window appears, enable averaging and set the # of
averages to 256 as shown below.
3 Filter out waveform noise by forcing the measurement calculations to wait until the
average is computed.
a Select Measure > Setup.
b Check the Wait for average to complete for measurements box.
c Click Close.
4 Set the Cal Out voltage (V
Cal Out
) to +30 mV as follows:
a Select Utilities > Calibration Output.
b Change the Signal Output function to DC (top left corner).
c Set the Level to 30 mV.
d Click Close.