Specifications

Chapter 3: Testing Performance
Offset Accuracy Test
19
b When the Acquisition Setup window appears, enable averaging and set the # of
averages to 256 as shown below.
3 Filter out waveform noise by forcing the measurement calculations to wait until the
average is computed.
a Select Measure > Setup.
b Check the Wait for average to complete for measurements box.
c Click Close.
4 Configure the scope to measure Average voltage on channel 1 as follows:
a Change the vertical scale of channel 1 to 10 mV/div.