Specifications
Chapter 2: Calibration
Running a self calibration
11
Running a self calibration
The self calibration uses signals generated in the oscilloscope to calibrate channel
sensitivity, offsets, and trigger parameters. There are three times we recommend
performing a self calibration:
• At least once a year.
• When you replace the hard drive.
• When the oscilloscope’s operating temperature (after the 30- minute warm- up period)
is more than ±5 °C different from that of the last calibration. Be sure to perform a
self calibration—even if one was recently performed—when environmental temperature
conditions cause the oscilloscope’s operating temperature to change, such as when
the oscilloscope is moved to a test rack or chamber.
Equipment Required
1 Let the oscilloscope warm up before running the self calibration.
The self calibration should be done only after the oscilloscope has run for 30 minutes
at ambient temperature with the cover installed. Calibration of an oscilloscope that has
not warmed up may result in an inaccurate calibration.
2 Select Utilities > Calibration.
3 Uncheck the Cal Memory Protect box.
You cannot run a self calibration if this box is checked. See Figure 2- 1.
Let the Oscilloscope Warm Up Before Adjusting
Warm up the oscilloscope for 30 minutes before starting the calibration procedure. Failure to allow warm
up may result in inaccurate calibration.
Equipment Critical Specifications Agilent Part Number
Connector savers (5 supplied
with oscilloscope)
3.5mm (f) to 3.5 mm (f) Agilent 5061-5311 (for 20, 25, 28, 30,
and 32 GHz models)
Agilent 1250-3758 (for 13 GHz
and16 GHz models)
Cable assembly 50 Ohm characteristic impedance BNC
(m) connectors, 36 inches (91 cm) to 48
inches (122 cm) long
Agilent 8120-1840
Adapter SMA(m) to BNC(f) Agilent 1250-1200
Cable (supplied with
oscilloscope)
No substitute Agilent 54916-61626
Digital channels cable (MSO
models only)
No substitute Agilent N2815-68701
MSO calibrator (MSO models
only)
No substitute Agilent N2834-68703
Calibration time
It will take approximately 2 hours and 15 minutes to run the self calibration on the oscilloscope, including
the time required to change cables from channel to channel.