Specifications
Chapter 5: Replacing Assemblies
To set the calibration factors after replacing the acquisition board
100
To set the calibration factors after replacing the acquisition
board
The calibration/self test procedure must be performed after replacing the acquisition
board. Consult the calibration chapter in this Service Guide.
Let the oscilloscope warm up before testing or calibrating.
The oscilloscope under test must be warmed up (with the oscilloscope application running) for at least
30 minutes prior to the start of any calibration or test.