Specifications
33
Hardware trigger
Sensitivity Internal low: 2.0 div p-p 0 to 22 GHz
Internal high: 0.3 div p-p 0 to 18 GHz, 1.0 div p-p 0 to 22 GHz
Edge trigger bandwidth >20 GHz
Minimum pulse width trigger
Hardware 250 ps
Software (InfiniiScan) 40 ps
Level range
Internal
Auxiliary
± 4 div from center screen or ± 4 Volts, whichever is smallest
± 5 V, also limit input signal to ± 5V
Sweep modes Auto, triggered, single
Display jitter
(displayed trigger jitter)
Equal to the TIE Jitter Measurement Floor (internal edge triggering with JitterFree)
Trigger sources Channel 1, Channel 2, Channel 3, Channel 4, aux, and line
Trigger modes
Edge
Edge transition
Edge then edge (time)
Edge then edge (Event)
Glitch
Pulse width
Triggers on a specified slope (rising, falling or alternating between rising and falling) and voltage
level on any channel or auxiliary trigger. Edge trigger bandwidth is > 20 GHz.
Trigger on rising or falling edges that cross two voltage levels in > or < the amount of time
specified. Edge transition setting from 250 ps.
The trigger is qualified by an edge. After a specified time delay between 10 ns to 10 s, a rising or
falling edge on any one selected input will generate the trigger
The trigger is qualified by an edge. After a specified delay between 1 to 16,000,000 rising or falling
edges, another rising or falling edge on any one selected input will generate the trigger.
Triggers on glitches narrower than the other pulses in your waveform by specifying a width less
than your narrowest pulse and a polarity. Triggers on glitches as narrow as 125 ps. Glitch range
settings: < 250 ps to < 10 s.
Trigger on a pulse that is wider or narrower than the other pulses in your waveform by specifying
a pulse width and a polarity. Triggers on pulse widths as narrow as 125 ps. Pulse width range
settings 250 ps to 10 s. Trigger point can be “end of pulse” or “time out”.
Runt Triggers on a pulse that crosses one threshold but fails to cross a second threshold before crossing
the first again. Can be time qualified with minimum setting of 250 ps.