Specifications

32
Acquisition 92004Q 92504Q 93304Q 95004Q 96204Q
Maximum real-time sample rate
(2 channels) 80 GSa/s 160 GSa/s 2ch
(4 Channels) 80 GSa/s 80 GSa/s
Memory depth per channel
4 channels (all models) 2 channels (all models)
Standard 20 Mpts 40 Mpts
Option 050M 50 Mpts (standard on DSA models) 100 Mpts (standard on DSA models)
Option 100 100 Mpts 200 Mpts
Option 200 200 Mpts 400 Mpts
Option 500 500 Mpts 1 Gpt
Option 01G 1 Gpt 1 Gpt
Option 02G 2 Gpts 2 Gpts
Maximum acquired time at highest real time resolution
Real-time resolution 80 GSa/s 160 GSa/s
Resolution 12.5 ps 6.25 ps
Standard (20M) 0.25 ms 0.125 ms
Option 50M 0.625 ms 0.3125 ms
Option 100 1.25 ms 0.625 ms
Option 200
Option 500
2.5 ms
6.25 ms
1.25 ms
3.125 ms
Option 01G 12.5 ms 6.25 ms
Option 02G 25 ms 12.5 ms
Sampling modes
Real-time Successive single shot acquisitions
Real-time with averaging Selectable from 2 to 65534 (up to 200,000 with function)
Real-time with peak detect 80 GSa/s (unavailable on RealEdge channels)
Real-time with hi resolution Real-time boxcar averaging reduces random noise and increases resolution (unavailable on RealEdge
channels)
Gaussian magnitude,
linear phase
Slower filter roll off while maintaining linear phase
Roll mode Scrolls sequential waveform points across the display in a right-to-left rolling motion. Works at sample
rates up to 10 MSa/s with a maximum record length of 40 Mpts
Segmented memory
5
Captures bursting signals at max sample rate without consuming memory during periods of inactivity
Number of segments (Up to 131,072 with >500M of memory depth)
Maximum time between triggers is 562,950 seconds
Re-arm time: 2.5 s
Maximum memory depth: Up to 8 Gpts in 1/2 channel mode with option 02G
Filters
Sin(x)/x Interpolation
On/off selectable FIR digital filter (2x, 4x, 8x 16x settings). Digital Signal Processing adds points between
acquired data points to enhance measurement accuracy and waveform display
Specifications
5. Segmented memory is currently not supported on 50 and 63 Ghz models of the 90000 Q-Series