Technical data

Table Of Contents
Trigger Commands 33
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 961
:TRIGger:SHOLd:HoldTIMe (HTIMe)
Command :TRIGger:SHOLd:HoldTIMe <time>
This command specifies the amount of hold time used to test for both a
setup and hold trigger violation. The hold time is the amount of time that
the data must be stable and valid after a clock edge.
<time> Hold time, in seconds.
Query
:TRIGger:SHOLD:HoldTIMe?
The query returns the currently defined hold time for the setup and hold
trigger violation.
Returned Format
[:TRIGger:SHOLD:HoldTIMe] <time><NL>