Technical data

Table Of Contents
956 Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference
33 Trigger Commands
:TRIGger:RUNT:TIME
Command :TRIGger:RUNT:TIME <time>
This command sets the time qualifier. The oscilloscope will trigger on a
runt pulse that has a width greater than the specified time.
<time> A real number for the time greater than qualifier, ranging from 250 ps to
30 ns.
Query
:TRIGger:RUNT:TIME?
The query returns the currently specified glitch width.
Returned Format
[:TRIGger:RUNT:TIME] <time><NL>