Technical data

Table Of Contents
938 Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference
33 Trigger Commands
:TRIGger:EDGE:SOURce
Command :TRIGger:EDGE:SOURce {CHANnel<N> | DIGital<M> | AUX | LINE }
The :TRIGger:EDGE:SOURce command selects the source for edge mode
triggering. This is the source that will be used for subsequent
:TRIGger:EDGE:SLOPe commands or queries.
<N> An integer, 1-4, for all other Infiniium Oscilloscope models.
<M> An integer, 0-15. The digital channels are only available on the MSO
oscilloscopes or DSO models with the MSO license installed.
Query
:TRIGger:EDGE:SOURce?
The query returns the currently selected edge mode trigger source.
Returned Format
[:TRIGger:EDGE:SOURce] {CHANnel<N> | DIGital<M> | AUX | LINE }<NL>