Technical data

Table Of Contents
Trigger Commands 33
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 907
:TRIGger:AND:SOURce
Command :TRIGger:AND:SOURce CHANnel<N>,{HIGH | LOW | DONTcare}
The :TRIGger:AND:SOURce command sets the logic value used to qualify
the trigger for the specified channel. The TRIGger:LEVel command
determines what voltage level is considered a HIGH or a LOW logic value.
If you set more than one channel to a HIGH or a LOW, then the multiple
channels are used to qualify the trigger.
<N> An integer, 1-4.
Query
:TRIGger:AND:SOURce? CHANnel<N>
The query returns the logic value for the designated channel.
Returned Format
[:TRIGger:AND:SOURce CHANnel<N>] {HIGH | LOW | DONTcare}<NL>