Technical data

Table Of Contents
818 Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference
28 Serial Bus Commands
:SBUS<N>:SPI:SOURce:DATA
Command :SBUS<N>:SPI:SOURce:DATA <source>
<source> ::= {CHANnel<N> | FUNCtion<F> | WMEMory<N> | DIGital<M>}
The :SBUS<N>:SPI:SOURce:DATA command sets the source for the SPI
serial MOSI data.
This command is the same as the :SBUS<N>:SPI:SOURce:MOSI command.
<N> CHANnel<N> is an integer, 1- 4.
WMEMory<N> is an integer, 1- 4.
<F> FUNCtion<F> is an integer, 1- 16.
<M> An integer, 0- 15. Digital channels are available on mixed-signal
oscilloscopes.
Query
:SBUS<N>:SPI:SOURce:DATA?
The :SBUS<N>:SPI:SOURce:DATA? query returns the current source for the
SPI serial MOSI data.
Returned Format
[:SBUS<N>:SPI:SOURce:DATA] <source><NL>
See Also ":SBUS<N>:SPI:SOURce:MOSI" on page 821
":SBUS<N>:SPI:SOURce:MISO" on page 820
":SBUS<N>:SPI:SOURce:CLOCk" on page 817
":SBUS<N>:SPI:SOURce:FRAMe" on page 819
":SBUS<N>:MODE" on page 792